White blood cell differential


A white blood cell differential is a medical laboratory test that provides information about the types and amounts of white blood cells in a person's blood. The test, which is usually ordered as part of a complete blood count, measures the amounts of the five normal white blood cell types – neutrophils, lymphocytes, monocytes, eosinophils and basophils – as well as abnormal cell types if they are present. These results are reported as percentages and absolute values, and then they are compared against reference ranges to determine whether the values are normal, low, or high. Changes in the amounts of white blood cells can aid in the diagnosis of many health conditions, including viral, bacterial, and parasitic infections and blood disorders such as leukemia.
White blood cell differentials may be performed by an automated analyzer – a machine designed to run laboratory tests – or manually, by examining blood smears under a microscope. The test was performed manually until white blood cell differential analyzers were introduced in the 1970s, making the automated differential possible. In the automated differential, a blood sample is loaded onto an analyzer, which samples a small volume of blood and measures various properties of white blood cells to produce a differential count. The manual differential, in which a medical laboratory scientist counts white blood cells on a stained microscope slide, is now performed to investigate abnormal or questionable results from the automated differential, or upon request by the healthcare provider. The manual differential can identify cell types that are not counted by automated methods and detect clinically significant changes in the appearance of white blood cells.
In 1675, Antonie van Leeuwenhoek published the first microscopic observations of blood cells, writing that blood consisted of "small red globules". Improvements in microscope technology throughout the 18th and 19th centuries allowed the cellular components of blood to be identified and counted. In the 1870s, Paul Ehrlich invented a staining technique that could differentiate between each type of white blood cell. Dmitri Leonidovich Romanowsky later modified Ehrlich's stain to produce a wider range of colours, creating the Romanowsky stain, which is still used to stain blood smears for manual differentials.
Automation of the white blood cell differential began with the invention of the Coulter counter, the first automated hematology analyzer, in the early 1950s. This machine used electrical impedance measurements to count cells and determine their sizes, allowing white and red blood cells to be enumerated. A cell sorting technique developed in 1965 related the sizes of white blood cells, as measured by Coulter counting, to cell types, enabling later methods which used cell size histograms to produce differential white blood cell counts. Throughout the 1970s, a number of methods were developed to improve cell identification, such as light scattering measurements and fluorescent staining. Hematology analyzers now use combinations of these and other methods to produce white blood cell differential results.

Medical uses

The white blood cell differential is a common blood test that is often ordered alongside a complete blood count. The test may be performed as part of a routine medical examination; to investigate certain symptoms, particularly those suggestive of infection or hematological disorders; or to monitor existing conditions, such as blood disorders and inflammatory diseases.
Five types of white blood cells are normally found in blood: neutrophils, lymphocytes, monocytes, eosinophils and basophils. Marked shifts in the proportions of these cell types, as measured by the automated or manual differential, can indicate various health conditions. Additionally, cell types which do not normally occur in the blood, such as blast cells, can be identified by the manual differential. These cell types may be found in blood disorders and other pathological states. The manual differential can also identify changes in the appearance of white blood cells, such as reactive lymphocytes, or features such as toxic granulation and vacuolation in neutrophils. The results of the white blood cell differential are reported as percentages and absolute values. Absolute counts are usually reported in units of cells per microliter or 109 cells per liter.
CBC and differential testing can be performed on venous or capillary blood. Capillary blood draws are generally used for infants and individuals whose veins are difficult to access. To prevent clotting, the sample is drawn into a tube containing the anticoagulant compound ethylenediaminetetraacetic acid. Tubes containing sodium citrate may also be used for patients in whom EDTA causes platelet clumping. The test is performed on whole blood, meaning blood that has not been centrifuged.

History

The first microscopic observations of blood cells were published by Antonie van Leeuwenhoek in 1675. Using a microscope of his own design, van Leeuwenhoek discovered that blood consisted of "small red globules, driven through a crystalline humidity of water". Throughout the 18th and 19th centuries, further improvements in microscope technology such as the invention of achromatic lenses allowed white blood cells and platelets to be counted in unstained blood samples. In the 1870s, Paul Ehrlich developed a staining technique that could differentiate between the five white blood cell types. Ehrlich's stain used a combination of an acidic and basic dye to stain white and red blood cells simultaneously. Dmitri Leonidovich Romanowsky improved on this technique in the 1890s by using a mixture of eosin and aged methylene blue, which produced a wide range of hues that was not present when either of the stains was used alone. This was termed the Romanowsky effect and became the basis for Romanowsky staining, the technique that is still used to stain blood smears for manual differentials.
The first automated hematology analyzer, the Coulter counter, was invented in the early 1950s by Wallace H. Coulter and Joseph A. Coulter. The analyzer worked on the Coulter principle, which states that when cells are suspended in a fluid carrying an electric current and passed through an aperture, they cause decreases in current proportional to their volume because of their poor electrical conductivity. The number and magnitude of these decreases can be used to count blood cells and calculate their sizes. The Coulter counter was initially designed for counting red blood cells, but it proved effective for counting white blood cells as well.
In 1967, M. J. Fulwyler used a modified version of the Coulter counter to sort white blood cells by size. This cell sorting method was based on a technology invented by R. G. Sweet for use in inkjet printers, which used electrostatic deflection to position ink droplets. Fulwyler's method used Coulter counting to determine the volume of the white blood cells, then partitioned cells into fluid droplets and applied charges based on the cells' sizes, allowing them to be sorted into collection vessels by volume. The collected cells could then undergo further analysis.
Using this method, Fulwyler found that human white blood cells formed three distinct peaks when their volumes were plotted on a histogram. By microscopically examining cells from each peak, he determined that the smallest-volume peak mainly contained lymphocytes, the intermediate-volume peak contained neutrophils, and the largest-volume peak contained a mixture of granulocytes and monocytes. Other researchers applied Fulwyler's techniques to generate automated neutrophil counts, although counts for other cell types were not as reliable. In 1983, the Coulter S Plus system, a commercial analyzer that produced a three-part differential of neutrophils, monocytes and lymphocytes based on cell size histograms, was introduced. The principle of identifying white blood cells by their sizes is still used in many hematology analyzers, although other techniques are used to improve discrimination between white blood cell types.
In the late 1960s and early 1970s, researchers began to combine Coulter counting techniques with analysis of the optical properties of cells. These devices, such as Kamentsky's Cytofluorograph and Dittrich & Goehde's Impulscytometer, shot beams of light at cells in specific wavelengths and measured the resulting absorbance or fluorescence, allowing cellular contents such as DNA to be quantified. Other researchers refined these techniques by using fluorescent dyes, which bind to specific components of cells to improve discrimination between cell types.
Meanwhile, other researchers such as George & Groner and Ansley & Ornstein focused on using light scattering patterns to classify cells. By directing a laser at individual cells and measuring the light scattered at different angles, various features of white blood cells can be ascertained; the light scattered within 1 to 10 degrees of the beam's axis describes cellular size, while light scattered at a 90 degree angle describes granularity. Ansley & Ornstein's work, along with later improvements by Mansfield, resulted in the first commercial flow cytometric differential analyzer, the Hemalog D. Introduced in 1974, this analyzer used light scattering and three different staining channels to identify the five normal white blood cell types and what were referred to as large unstained cells, a classification that usually consisted of atypical lymphocytes or blast cells. The Hemalog D could count 10,000 cells in one run, a massive improvement over the 100-cell manual differential, and it decreased the workload in the hematology laboratory as samples that were identified as normal did not require a manual smear review.
In 1981, Technicon combined the Hemalog D with the Hemalog-8 CBC analyzer to produce the Technicon H6000, the first combined CBC and differential analyzer. This analyzer was unpopular with hematology laboratories because it was labour-intensive to operate. However, in the late 1980s to early 1990s similar systems were widely produced by other manufacturers such as Sysmex, Abbott, Roche and Beckman Coulter.

Types

Manual differential

In a manual differential, a medical laboratory scientist examines a stained blood smear under a microscope and counts and classifies white blood cells based on their appearance. A manual differential is usually performed when the automated differential is flagged for review or when the healthcare provider requests it. If the manual differential shows findings suggestive of certain serious conditions, such as leukemia, the blood smear is referred to a physician for confirmation.

Procedure

A blood smear is prepared by placing a drop of blood on a microscope slide and using a second slide held at a 30 to 45 degree angle to spread the blood and pull it across the slide, forming a "feathered edge" consisting of a single layer of cells at the end of the smear. This may be done by hand or using an automated slide maker coupled to a hematology analyzer. The slide is treated with a Romanowsky stain, commonly Wright's stain or Wright-Giemsa, and examined by a medical laboratory scientist. The smear is examined in a systematic pattern, scanning from side to side within the feathered edge and counting cells consecutively. The differential is typically performed at 400x or 500x magnification, but 1000x magnification may be used if abnormal cells are present. Cells are identified based on their morphologic features, such as the size and structure of their nucleus and the colour and texture of their cytoplasm. This allows abnormal cell types and changes in cellular appearance to be identified. In most cases, the scientist counts 100 white blood cells, but 200 may be counted for better representation if the white blood cell count is high. The manual differential count produces percentages of each cell type, which can be multiplied by the total white blood cell count from the analyzer to derive the absolute values.
The manual differential can be partially automated with digital microscopy software, which uses artificial intelligence to classify white blood cells from photomicrographs of the blood smear. However, this technique requires confirmation by manual review.

Limitations

Because relatively few cells are counted in the manual differential, the variability is higher than in automated techniques, especially when cells are present in low amounts. For example, in a sample containing 5 percent monocytes, the manual differential results could be between 1 and 10 percent due to sampling variation. Additionally, cell identification is subjective and the accuracy depends on the skills of the person reading the slide. Poor blood smear preparation can cause an uneven distribution of white blood cells, resulting in inaccurate counting, and improper staining can impede cell identification. Overall, manual differential counts exhibit coefficients of variation ranging from 5 to 10 percent, while automated differential counts of normal neutrophils and lymphocytes have CVs of about 3 percent.
In leukemias and other hematologic malignancies, the lineage and genetic characteristics of white blood cells have important implications for treatment and prognosis, and the microscopic appearance of the cells is often insufficient for accurate classification. In these cases, other techniques such as immunophenotyping by flow cytometry or special staining can be used to definitively identify the cells.

Automated differential

Most hematology analyzers provide a five-part differential, enumerating neutrophils, lymphocytes, monocytes, eosinophils and basophils. Some instruments can also count immature granulocytes and nucleated red blood cells. Hematology analyzers measure various properties of white blood cells, such as impedance, light scattering parameters, and staining reactions. This data is analyzed and plotted on a scattergram, forming distinct clusters which correspond to white blood cell types. The number of points in each cluster describes the number of cells, and the shape of the clusters can indicate changes in cellular structure. The analyzer counts many more cells than are counted in a manual differential, resulting in improved precision. If abnormal features or cell populations that the analyzer cannot identify are present, the instrument can flag the results for manual blood smear review.

Procedure

Most hematology analyzers use combinations of light scattering, Coulter counting, and cytochemical staining techniques. Some analyzers also use radiofrequency analysis and monoclonal antibody tagging to identify cells. Staining techniques used in differential analyzers include staining of myeloperoxidase, an enzyme found in cells of myeloid lineage, and nucleic acids, which are found in higher concentrations in immature cells.
A small volume of blood is aspirated into the analyzer, where reagents are applied to lyse red blood cells and preserve white blood cells. The sample is diluted and passed into a flow cell, which uses hydrodynamic focusing to isolate single cells for accurate analysis of their properties. Various cellular parameters, such as size, complexity and staining reactions, are measured and analyzed to identify cell populations. Basophils, which are difficult to differentiate from other white blood cells by routine methods, are usually quantified using a reagent that lyses the cytoplasm of other white blood cells but leaves basophils intact. Samples that have abnormal results or are suspected to contain abnormal cells are flagged by the analyzer for manual blood smear review.
To ensure that results from the automated analyzer are correct, quality control samples are run at least once per day. These are samples with known results that are most often provided by the instrument manufacturer. Laboratories compare their differential results to the known values to ensure the instrument is operating correctly. A moving average measurement may also be used, in which the average results for patient samples are measured at certain intervals. Assuming that the characteristics of the patient population remain roughly the same over time, the average should remain constant. Large shifts in the average value can indicate instrument problems.

Limitations

When immature or abnormal white blood cells are present, automated differential results may be incorrect, necessitating a manual blood smear review. Overall, 10 to 25 percent of CBC samples are flagged for manual review by the analyzer. While the majority of abnormal samples are automatically flagged, some are missed; conversely, analyzers may generate false positive flags when no abnormal cells are present. Hematology laboratories compensate for these issues by requiring a smear review when differential or CBC results fall outside certain numerical thresholds, regardless of the presence of analyzer flags. The sensitivity and specificity of analyzer flagging can be determined by comparing analyzer flags to manual differential results.
The automated basophil count is notoriously unreliable, often underestimating counts in basophilia and producing falsely elevated results in the presence of abnormal cells. The manual differential is therefore considered the reference method for these cells.
Analyzers may count nucleated red blood cells, giant and clumped platelets, and red blood cells containing abnormal hemoglobins as white blood cells, leading to faulty differential results. Automated differential counts on aged specimens may be incorrect due to cellular degeneration.

White blood cell types and result interpretation

s for the white blood cell differential are defined by individual laboratories and may vary due to different patient populations and testing methods. The ranges presented here are for example purposes only.
Cell typeImageDescriptionIncrease Decrease Adult reference range Adult reference range
NeutrophilNeutrophils are the most common white blood cells in normal adult blood. When stained with a Romanowsky stain, they exhibit a multi-lobed nucleus and pink cytoplasm that contains small purple granules.Neutrophilia: Acute infections, especially bacterial; inflammation, physiologic stress, myeloproliferative disorder; normal in pregnancy and the neonatal periodNeutropenia: Various drugs, especially chemotherapy; leukemia, myelodysplastic syndrome and other bone marrow disorders; certain infections; autoimmune and congenital disorders; may be normal in certain ethnicities50–701.7–7.5
LymphocyteLymphocytes, which are the second most common type of white blood cell in adults, are typically small cells with a round, dark nucleus and a thin strip of pale blue cytoplasm. Some lymphocytes are larger and contain a few blue granules.Lymphocytosis: Viral infection, chronic lymphocytic leukemia and other lymphoproliferative disorders, splenectomy; normal in childrenLymphopenia: HIV/AIDS, influenza, viral hepatitis; protein-energy malnutrition, drug reaction, various acute illnesses18–421.0–3.2
MonocyteMonocytes are large cells with a curved or folded nucleus and finely granulated, grey-blue cytoplasm that often contains vacuoles. Monocytes are the third most common white blood cell after neutrophils and lymphocytes.Monocytosis: Chronic infection and inflammation, myeloproliferative disorders and monocytic leukemias; normal in newbornsMonocytopenia: Chemotherapy, aplastic anemia, severe burns, AIDS2–110.1–1.3
EosinophilEosinophils have large orange granules in their cytoplasm and bi-lobed nuclei. They are found in low amounts in normal blood.Eosinophilia: Parasitic infection, allergic conditions, asthma, myeloproliferative disorderEosinopenia: Normally found in low numbers; may be decreased during pregnancy, acute stress and inflammation, and treatment with certain drugs1–30–0.3
BasophilBasophils exhibit large, dark purple granules that often cover the cell's nucleus. They are the rarest of the five normal cell types.Basophilia: Myeloproliferative disorder, splenectomy, hypersensitivity reactionsBasopenia: Normally found in low numbers; may be decreased during acute stress, ovulation and treatment with steroid drugs0–20–0.2
Band neutrophilBand neutrophils are young forms of neutrophils which lack segmentation of the nucleus. These cells, which are identified by manual counting, are found in low numbers in normal adult blood. The term left shift refers to an increase in bands or immature granulocytes.Left shift or bandemia: Infection, inflammation, myeloproliferative disorder; normal in pregnancyN/AN/AN/A
Immature granulocyteImmature granulocytes are immature forms of neutrophils and other granulocytes. This classification consists of metamyelocytes, myelocytes and promyelocytes, which may be enumerated separately in the manual differential or reported together as immature granulocytes by automated methods. Immature granulocytes are normally found in the bone marrow, but not in peripheral blood.Left shift: Infection, inflammation, myeloproliferative disorder, steroid use, pregnancyNot normally found in peripheral bloodN/AN/A
Blast cellBlast cells are very immature cells that are normally found in the bone marrow, where they develop into mature cells before being released into the blood. They can be identified by their large overall size, deep blue cytoplasm, and large nucleus with fine chromatin and prominent nucleoli.Leukemia and other hematologic disorders; may rarely be seen in severe cases of left shiftNot normally found in peripheral bloodN/AN/A

Clinical significance

; Neutrophil : The neutrophil count is normally higher in newborns and pregnant women than in other groups. Outside of these conditions, increased neutrophil counts are associated with bacterial infection, inflammation, and various forms of physiological stress. Neutrophil counts can become extremely high in some infections and inflammatory states, which is termed leukemoid reaction because the high white blood cell count mimics leukemia. Neutrophilia may also occur in myeloproliferative disorders.
Neutropenia, meaning a low neutrophil count, may occur as a response to drug treatment or in certain infections, such as tuberculosis and Gram-negative sepsis. Neutropenia also occurs in many hematologic disorders, such as leukemia and myelodysplastic syndrome, and in a variety of autoimmune and congenital diseases. A decreased neutrophil count may be normal in individuals of certain ethnicities; this is termed benign ethnic neutropenia. Very low neutrophil counts are associated with immunosuppression.
When stimulated by infection or inflammation, neutrophils may develop abnormal features in their cytoplasm, such as toxic granulation, toxic vacuolation and Döhle bodies. These features, which are caused by the release of cytokines, are collectively known as toxic changes.
; Lymphocyte : Increased lymphocyte counts can be caused by various viral infections and may also occur after splenectomy. Children have higher lymphocyte counts than adults. Chronic lymphocytic leukemia presents with an elevated lymphocyte count and abnormal lymphocyte morphology, in which the lymphocytes have extremely dense, clumped nuclei and some cells appear smudged on the blood smear.
Low lymphocyte counts may be seen in infections such as HIV/AIDS, influenza and viral hepatitis, as well as in protein-energy malnutrition, acute illnesses and drug reactions.
In response to viral infections, lymphocytes may increase greatly in size, developing unusually shaped nuclei and large amounts of dark blue cytoplasm. Such cells are referred to as reactive or atypical lymphocytes and when present they are either commented on or counted separately from normal lymphocytes in the manual differential.
;Monocyte: Increased monocyte counts are seen in chronic infection and inflammation. Extremely high monocyte counts, as well as immature forms of monocytes, occur in chronic myelomonocytic leukemia and acute leukemias of monocytic origin. Monocyte counts may be decreased in individuals who are receiving chemotherapy as well as those with aplastic anemia, severe burns, and AIDS.
;Eosinophil: Elevated eosinophil counts are associated with allergic reactions, parasitic infections, and asthma. Eosinophil counts may be decreased in pregnancy and in response to physiological stress, inflammation or treatment with certain drugs, such as steroids and epinephrine.
;Basophil:Basophilia and eosinophilia can occur along with other white blood cell abnormalities in chronic myeloid leukemia and other myeloproliferative disorders. An increased basophil count may also be seen in hypersensitivity reactions and after splenectomy. The basophil count may decrease during ovulation, steroid treatment, and periods of physiological stress.
;Band neutrophil: A left shift, meaning an increase in band neutrophils or immature granulocytes, can indicate infection, inflammation or bone marrow disorders, although it can also be a normal finding in pregnancy. Some laboratories do not separate bands from mature neutrophils in the differential count because the classification is highly subjective and unreliable.
;Immature granulocyte: When present in significant quantities in the blood, immature granulocytes can indicate infection and inflammation, as well as myeloproliferative disease, leukemia and other conditions affecting the marrow. IGs may also be increased in steroid use and pregnancy. Chronic myeloid leukemia often presents with a high number of immature granulocytes in the peripheral blood. Abnormal promyelocytes with multiple Auer rods, called faggot cells, occur in acute promyelocytic leukemia.
;Blast cell: When seen on the blood smear, blast cells are an abnormal finding and may be indicative of acute leukemia or other serious blood disorders. Rarely, they may be seen in severe cases of left shift. The presence of Auer rods inside blast cells indicates that they are of myeloid origin, which has important implications for leukemia treatment. Other morphologic features can provide information about the lineage of blast cells: for example, myeloblasts tend to be large with distinct nucleoli, while lymphoblasts can be smaller with a denser chromatin pattern. However, these features are not diagnostic, and flow cytometry or special staining is generally used to confirm the lineage.
;Others: Various other abnormal cells may be present in the blood in certain conditions. For example, lymphoma cells may be found on the manual differential in some cases of lymphoma, and in mast cell leukemia, mast cells, which are normally confined to tissue, circulate in the blood. There is a very rare phenomenon called carcinocythemia in which tumour cells are seen on the peripheral blood smear.