Since the piston makes direct contact with the sample, the aspiration force in a positive displacement pipette is unaffected by the sample's physical properties. Several liquid handling companies suggest that positive displacement pipettes can be used to accurately pipette very viscous, volatile, hot or cold, or corrosive samples.
, such as glycerol, flow very slowly. Glycerol has high dynamic viscosity, and if a researcher aspirates a sample of glycerol too quickly with an air displacement pipette, It will draw up air bubble. When a researcher attempts to dispense the liquid, some of it will stick to the pipette tip wall, dispense very slowly and remain in the tip. Surfactants also produce this effect, but the remaining liquid film is thinner. In a positive displacement pipette, the aspiration strength remains constant, so the tip fills evenly. Also, the piston slides along the internal sides of the pipette tip and pushes the total volume out, so no liquid is left behind.
Volatile liquids
such as acetone, hexane, and methanol, evaporate continuously in air displacement pipettes. Some volatile liquids expand so quickly that they expand the air column in the pipette, which causes leakage: The pipette will lose drops and dispense liquid imprecisely. As drops leak out, they can contaminate the bench, ultimately causing cross-contamination from sample to sample. These drops can also produce a health hazard. Because there is no air cushion in a positive displacement pipette, liquids do not evaporate or leak. Drops will not fall from the tip, and vapors will not contaminate the internal parts of the pipette. Also, the capillary/piston tips used for positive displacement pipetting are disposable.
In an air displacement pipette, the ambient temperature is correlated with the volume of the air cushion and affects the aspiration volume. Cold liquids, such as a suspension of restriction enzymes, which are usually handled at 0°C, cause the air cushion to shrink and the pipette to aspirate more liquid than expected, making the pipette over-deliver. Hot samples, such as mammaliancell cultures at body temperature or polymerase chain reaction solutions at 60°C or higher, will cause the air cushion to expand, causing the pipette to aspirate less liquid than expected and making the pipette under-deliver. Positive displacement pipettes do not have an air cushion and are less affected by liquid temperature, yielding greater pipetting accuracy.
Corrosive and hazardous liquids
and radioactive liquids may damage the piston, seal, and tip holder in an air displacement pipette. Positive displacement pipettes use a disposable capillary/piston tip, so the pipette is not affected by corrosive samples over its lifetime. Since there is no contact between the sample and the pipette, there is little risk of contamination.
Pipetting technique
Positive displacement pipettes operate very similarly to air displacement pipettes. Steps for operating a positive displacement pipette